Analysis of Tagatose According to USP-NF Method (SC1011)

According to the United States Pharmacopeia and the National Formulary (USP 40-NF 35*), the assay of tagatose should be carried out with a column with L22 packing material and meets following requirements. The SUGAR SC1011 confirmed the requirements were met.

System suitability requirements:
    Relative standard deviation (RSD): ≤ 2.0 %

*The version at the time of the application acquisition.


Sample : 20 µL
1. Tagatose 5 mg/mL


Column       : Shodex SUGAR SC1011 (8.0 mm I.D. x 300 mm)
Eluent       : 0.05 mg/mL Calcium acetate aq.
Flow rate    : 0.6 mL/min
Detector     : RI
Column temp. : 85 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)