According to the United States Pharmacopeia and the National Formulary (USP 42-NF 37*), analysis of maltose should be carried out with a column with L58 packing material and meets following requirements. The SUGAR KS-801 confirmed the requirements were met.
System suitability requirements:
Resolution of maltose and maltotriose: ≥ 1.6
Relative standard deviation (RSD) of maltose peak area: ≤ 2.0 %
*The version at the time of the application acquisition.
Sample : 10 mg/mL each
(in H2O), 20 µL
1. Maltotriose
2. Maltose
3. Glucose
Column : Shodex SUGAR KS-801 (8.0 mm I.D. x 300 mm) Eluent : H2O Flow rate : 0.35 mL/min Detector : RI Column temp. : 80 °C
Please select your location to continue:
For all other locations please click “close” and return to the homepage.
This website stores cookies on your computer. These cookies are used to collect information about how you interact with our website and allow us to remember you. We use this information in order to improve our services. If you are happy to accept these cookies please click “Accept” or simply continue to use our website. To find out more about the cookies we use, see “More Information”.