Analysis of Dextrose According to USP-NF Method (SC1011)

According to the United States Pharmacopeia and the National Formulary (USP 40-NF 35*), "ASSAY" and "IMPURITIES: Related Substances" for "Dextrose" should be carried out with a column filled with L19 packing material. The SUGAR SC1011, a column filled with L19 packing material, confirmed the requirement was met.

System suitability requirementResolution between maltotriose and maltose (System suitability solution): ≥ 1.3

*The version at the time of the application acquisition.

Sample: 20 µL
(System suitability solution) 0.1 mg/mL each of Maltose, Maltotriose and Fructose in H2O
(Standard solution A) 30 mg/mL of Dextrose in H2O

  1. 1.Maltotriose
  2. 2.Maltose
  3. 3.Dextrose
  4. 4.Fructose
Column
Shodex SUGAR SC1011 (8.0 mm I.D. x 300 mm)
Eluent
H2O
Flow rate
0.3 mL/min
Detector
RI (40 ℃)
Column temp.
85 ℃