LC/MS Analysis of Metformin (NH2P-40 2D)

High polar compounds, such as metformin (an oral antidiabetic drug in biguanide class) are not retained by reversed-phase chromatography. Instead, HILIC mode is suitable for the analysis of polar compounds. A polymer-based amino column, Asahipak NH2P-40 2D, was used in the LC/MS method. Eluent containing a high acetonitrile ratio provides a high sensitive MS detection.
*Sales of Asahipak NH2P-40 2D used in this application is discontinued. Please consider trying Asahipak NH2P-50 2D.

Sample: 5 μL

  1. 1.Metformin-HCl 5 ng/mL (in H2O/CH3CN=5/95)
Column
Shodex Asahipak NH2P-40 2D (2.0 mm I.D. x 150 mm)
Eluent
50 mM HCOONH4 aq./CH3CN=5/95
Flow rate
0.2 mL/min
Detector
ESI-MS (SIM)
Column temp.
30 ℃

Sample Name Index