Comparison of HK-404L and KF-805L

Poly(isobutyl methacrylate) was analyzed using GPC HK-404L, an organic SEC (GPC) column for ultra-rapid analysis, and  KF-805L, a conventional SEC column. The elution volume per column of HK-404L is about 2.0 mL. The column pressure of HK-404L is low at high flow rate and the eluent can pass through the column at the same flow rate as the conventional column with the conventional HPLC instrument. Therefore the analytical time is decreased to 1/6 of conventional analysis and ultra-rapid analysis becomes possible.


Sample : 5 µL
Poly(isobutyl methacrylate)
0.2 %


Columns      : Shodex GPC HK-404L (4.6 mm I.D. x 150 mm)
               Shodex GPC KF-805L (8.0 mm I.D. x 300 mm)
Eluent       : THF
Flow rate    : 1.0 mL/min
Detector     : RI (small cell volume)
Column temp. : 40 °C