The effects of sample concentration on separation of EPON1001 polymer were examined using organic solvent SEC (GPC) semi-micro columns, GPCKF-403HQ and KF-402HQ. No considerable decline in resolution was observed up to analysis of 5 % (m/v) EPON1001.
Please refer to Effects of Sample Concentration (1) (KF-402HQ) for the analysis of sample with single component.
Sample : EPON 1011, 10 μL
Columns : Shodex GPC KF-403HQ + KF-402HQ (4.6 mm I.D. x 250 mm each) Eluent : THF Flow rate : 0.3 mL/min Column temp. : 40 °C
Please select your location to continue:
For all other locations please click “close” and return to the homepage.
This website stores cookies on your computer. These cookies are used to collect information about how you interact with our website and allow us to remember you. We use this information in order to improve our services. If you are happy to accept these cookies please click “Accept” or simply continue to use our website. To find out more about the cookies we use, see “More Information”.