Effects of Sample Concentration (2) (KF-403HQ + KF-402HQ)

The effects of sample concentration on separation of EPON1001 polymer were examined using organic solvent SEC (GPC) semi-micro columns, GPCKF-403HQ and KF-402HQ. No considerable decline in resolution was observed up to analysis of 5 % (m/v) EPON1001.
Please refer to Effects of Sample Concentration (1) (KF-402HQ) for the analysis of sample with single component.


Sample : EPON 1011, 10 μL



Columns      : Shodex GPC KF-403HQ + KF-402HQ (4.6 mm I.D. x 250 mm each)           
Eluent       : THF       
Flow rate    : 0.3 mL/min
Column temp. : 40 °C