Comparison of LF-804, LF-404 and LF-604

GPC LF-804 is an Organic SEC (GPC) column packed with unique multi-pore gels having a wide pore-size distribution. GPC LF-404, a semi-micro column, and GPC LF-604, a downsized column, can reduce solvent use to less than one-third compared to that of LF-804, a conventional column. In addition, LF-404 offers a resolution comparable to LF-804. Meanwhile, LF-604 is suitable for rapid analysis and can complete analyses in about half the time required by LF-804.

Sample: Polystyrene standards

  1. 1.Mw 7,290,000
  2. 2.Mw 1,460,000
  3. 3.Mw 501,000
  4. 4.Mw 185,000
  5. 5.Mw 68,900
  6. 6.Mw 30,300
  7. 7.Mw 9,890
  8. 8.Mw 2,350
  9. 9.Mw 580
Column
Shodex GPC LF-804 (8.0 mm I.D. x 300 mm) x 2
Shodex GPC LF-404 (4.6 mm I.D. x 250 mm) x 2
Shodex GPC LF-604 (6.0 mm I.D. x 150 mm) x 2
Eluent
THF
Flow rate
(LF-804); 1.0 mL/min, (LF-404); 0.3 mL/min, (LF-604); 0.6 mL/min
Detector
RI (LF-804; conventional), (LF-404 and LF-604; small cell volume)
Column temp.
40 ℃