Polystyrene (Mp 580) was analyzed using GPC HK-401, an organic SEC (GPC) column for ultra-rapid analysis. Different number of HK-401 columns (1, 2, and 3 units) were used alone/together in series to compare their separation efficiencies. Use of multiple columns improves the separation of oligomers, although it also increases the analysis time. Regardless, the analysis time using three HK-401 columns at the flow rate of 1.0 mL/min takes less than 10 minutes.
Sample : 0.5 %, 5 µL
1. Polystyrene (Mp: 580)
2. Propylbenzene
Columns : Shodex GPC HK-401 (4.6 mm I.D. x 150 mm) x n Eluent : THF Flow rate : 1.0 mL/min Detector : RI (small cell volume) Column temp. : 30 °C
Please select your location to continue:
For all other locations please click “close” and return to the homepage.
This website stores cookies on your computer. These cookies are used to collect information about how you interact with our website and allow us to remember you. We use this information in order to improve our services. If you are happy to accept these cookies please click “Accept” or simply continue to use our website. To find out more about the cookies we use, see “More Information”.