GPC HK-400 series, organic SEC (GPC) columns for ultra-rapid analysis have low column pressure even at high flow rates and can be used in conventional HPLC device. The separation of epoxy resin (jER1001) using HK-402 was compared at flow rates of 0.3 mL/min and 1.0 mL/min. At a flow rate of 1.0 mL/min, equivalent separation can be achieved with an analysis time of 1/6 or less of a conventional SEC column, but if separation is important, it is recommended to analyze at 0.3 mL/min. (Even at 0.3 mL/min, the analysis time is less than half that of a conventional SEC column.)
Note : Epoxy Resin (6) (Comparison between HK-402 and KF-802.5)
Sample : 10 µL
jER1001 (Epoxy resin) 0.2 %(w/v)
Column : Shodex GPC HK-402 (4.6 mm I.D. x 150 mm) Eluent : THF Flow rate : 0.3 mL/min, 1.0 mL/min Detector : UV (254 nm) (small cell volume) Column temp. : 40 ℃
Please select your location to continue:
For all other locations please click “close” and return to the homepage.
This website stores cookies on your computer. These cookies are used to collect information about how you interact with our website and allow us to remember you. We use this information in order to improve our services. If you are happy to accept these cookies please click “Accept” or simply continue to use our website. To find out more about the cookies we use, see “More Information”.