Please use this column with suppressor type ion chromatography system.
According to Japanese Testing Methods for Fertilizers (2019*), "6.2.a Ion chromatography method" under "6.2 Chloride" includes both suppressor and non-suppressor methods. The suppressor method should be carried out with a column filled with porous polyvinyl alcohol gel modified with quaternary ammonium groups. The method mentions that this analysis is applicable to potassium sulfate, potassium bicarbonate, magnesium-containing potassium sulfate, fish meal powder, fish meal, and compost, and IC SI-52 4E as a suitable column. In this application, a standard solution and one target fertilizer, potassium sulfate, were analyzed using SI-52 4E. The calibration curve for chloride ion showed high linearity with a coefficient of determination (R2 = 0.9973) over the concentration range of 1 to 20 µg/mL.
Analysis of chloride in fertilizers by a non-suppressor method is also available here.
Sample Preparation according to "(4.1) Extraction"
- (1)Weigh 1.00 g of the test material and place it in a 200 mL screw-capped Erlenmeyer flask.
- (2)Add 100 mL of water and stir for 10 minutes on a shaker.
- (3)After standing, transfer the supernatant into a 50 mL screw-capped centrifuge tube and centrifuge at 1700 x g for about 5 minutes.
- (4)Dilute the supernatant precisely 20-fold with water.
- (5)Filter through a 0.45-µm membrane filter and use it as a sample.
*The version at the time of the application acquisition.
Sample: 20 µL each
(Standard solution) 5 mg/L of Cl-
- 1.Cl-
(Sample solution)
- 1.Cl-


- Column
- :Shodex IC SI-92G (4.0 mm I.D. x 10 mm) + SI-52 4E (4.0 mm I.D. x 250 mm)
- Eluent
- :6.4 mM Na2CO3 aq.
- Flow rate
- :0.8 mL/min
- Detector
- :Suppressed conductivity
- Column temp.
- :40 ℃
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Table of Anion Retention Time under Various Eluent Conditions (SI-90 4E)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-90 4E)
- Effects of Sodium Bicarbonate Concentration on Retention Time (SI-90 4E)
- Effects of Temperature on Elution Time (SI-90 4E)
- Effects of Added Acetonitrile on the Retention Time (SI-90 4E)
- Effects of Column Temperature on Retention Time (SI-50 4E)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-50 4E)
- Anion Standards (19) Rapid Analysis (SI-90 4E)
- Anion Standards (20) (Simultaneous Analysis of HS-, SO32-, SO42- and S2O32-) (SI-90 4E)
- Anion Standards (21) (Methanesulfonic Acid)
- Anion Standards (22) Rapid Analysis (SI-50 4D)
- Effects of Flow Rate (SI-50 4D)
- Effects of Column Temperature (SI-50 4D)
- High Sensitive Analysis of Anions (SI-90 4E)
- Anion Standards and Organic Acids (2) (SI-90 4E)
- Anion Standards and Organic Acids (3) (SI-50 4E)
- Mineral Water (2) (SI-90 4E)
- Sulfite in Wine (2) (SI-90 4E)
- Sodium Azide (2) (SI-90 4E)
- Sulfamic Acid (SI-90 4E)
- Sea Water (3) (SI-90 4E)
- Perchloric Acid (SI-90 4E)
- Analysis of Iodide Ion (SI-90 4E)
- Fluoroacetid Acids (SI-50 4E)
- Anion Analysis in Lithium Salts (3) (SI-90 4E)
- Analysis of Oxyhalides (SI-52 4E)
- Effects of Temperature on Retention Time (SI-52 4E)
- Calibration Curves for Oxyhalides (SI-52 4E)
- Analysis of Organic Acids and Oxyhalides (SI-52 4E)
- Analysis of Oxyhalides and Anions (SI-52 4E)
- Simultaneous Analysis of Oxyhalides and Anions in River Water (SI-52 4E)
- Simultaneous Analysis of Oxyhalides and Anions in Rain Water (SI-52 4E)
- Simultaneous Analysis of Oxyhalides and Anions According to Japanese Water Supply Act (SI-52 4E)
- Iodic Acid (SI-52 4E)
- Separation of Phosphate, Phosphtite and Hypophosphite (SI-52 4E)
- Separation of Phosphate, Phosphtite and Hypophosphite (SI-90 4E)
- Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (1)
- Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (2)
- Effects of Temperature on Retention Time (SI-35 4D)
- Simultaneous Analysis of Oxyhalides and Anions According to Japanese Water Supply Act (2) (SI-35 4D)
- Analysis of Organic Acids and Oxyhalides (SI-35 4D)
- Methanesulfonic Acid and Methanedisulfonic Acid (SI-35 4D)
- Calibration Curves (SI-35 4D)
- Effects of Injection Volume (SI-35 4D)
- Effects of Sample Concentration on Theoretical Plate Number (SI-35 4D)
- Analysis of Tricarboxylic Acids (SI-35 4D)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-35 4D)
- Analysis of Sulfite in Disposable Wooden Chopsticks (SI-35 4D)
- Analysis of Alkyl Sulfonic Acids (2) (SI-35 4D)
- Rapid Analysis of Anion Standards (SI-35 2B)
- Calibration Curves (SI-35 2B)
- Effects of Temperature on Retention Time (SI-35 2B)
- Analysis of Anions (SI-36 4D)
- Effects of Potassium Hydroxide Concentration on Retention Time (SI-36 4D)
- Effects of Potassium Hydroxide Concentration on Retention Time (2) (SI-36 4D)
- Effects of Temperature on Retention Time (SI-36 4D)
- Simultaneous Analysis of Common Oxyhalides and Anions (SI-37 4D)
- Analysis of Oxyhalides in Artificial-Drinking Water According to EPA Method 300.1 (SI-37 4D)