Please use this column with suppressor type ion chromatography system.
The Japanese regulation for “Testing Methods for Fertilizers (2019*)” lists SI-52 4E as a suppressor type IC (anion) column for the analysis of chloride in fertilizers. This method applies to fertilizers, such as “potassium sulfate”, “potassium bicarbonate”, “magnesium potassium sulfate”, “fish powder”, “fish meal”, and “compost”.
In this application, we used SI-52 4E to analyze chloride standard solution and potassium sulfate, one of the fertilizers included in the method. A good linearity was obtained for the calibration curve in the concentration range 1 to 20 μg/mL chloride.
Analysis of chloride in fertilizer using a non-suppressor method is also available here.
*The version at the time of the application acquisition.
Sample preparation
1. Weigh 1.00-g of the target sample and put it in a 200-mL Erlenmeyer flask with stopper.
2. Add 100-mL water and mix the contents for 10 minutes by placing the flask on a flask shaker.
3. Pour the supernatant into a 50-mL centrifuge tube with stopper and centrifuge at 1,700 x g for about 5 minutes.
4. Dilute the supernatant 1 + 19 with water.
5. Filter the mixture with a 0.45-µm membrane filter and use it as a sample.
Sample : 20 μL
Srandard 5 mg/L
Extracted sample
1. Cl-
Columns : Shodex IC SI-92G (4.0 mm I.D. x 10 mm) + SI-52 4E (4.0 mm I.D. x 250 mm) Eluent : 6.4 mM Na2CO3 aq. Flow rate : 0.8 mL/min Detector : Suppressed conductivity Column temp. : 40 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Table of Anion Retention Time under Various Eluent Conditions (SI-90 4E)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-90 4E)
- Effects of Sodium Bicarbonate Concentration on Retention Time (SI-90 4E)
- Effects of Temperature on Elution Time (SI-90 4E)
- Effects of Added Acetonitrile on the Retention Time (SI-90 4E)
- Effects of Column Temperature on Retention Time (SI-50 4E)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-50 4E)
- Anion Standards (19) Rapid Analysis (SI-90 4E)
- Anion Standards (20) (Simultaneous Analysis of HS-, SO32-, SO42- and S2O32-) (SI-90 4E)
- Anion Standards (21) (Methanesulfonic Acid)
- High Sensitive Analysis of Anions (SI-90 4E)
- Anion Standards and Organic Acids (2) (SI-90 4E)
- Anion Standards and Organic Acids (3) (SI-50 4E)
- Mineral Water (2) (SI-90 4E)
- Sulfite in Wine (2) (SI-90 4E)
- Sodium Azide (2) (SI-90 4E)
- Sulfamic Acid (SI-90 4E)
- Sea Water (3) (SI-90 4E)
- Perchloric Acid (SI-90 4E)
- Analysis of Iodide Ion (SI-90 4E)
- Fluoroacetid Acids (SI-50 4E)
- Anion Analysis in Lithium Salts (3) (SI-90 4E)
- Analysis of Oxyhalides (SI-52 4E)
- Effects of Temperature on Retention Time (SI-52 4E)
- Calibration Curves for Oxyhalides (SI-52 4E)
- Analysis of Organic Acids and Oxyhalides (SI-52 4E)
- Analysis of Oxyhalides and Anions (SI-52 4E)
- Analysis of Oxyhalides and Anions in River Water (SI-52 4E)
- Analysis of Oxyhalides and Anions in Rain Water (SI-52 4E)
- Analysis of Oxyhalides and Anions in Tap Water (SI-52 4E)
- Iodic Acid (SI-52 4E)
- Separation of Phosphate, Phosphtite and Hypophosphite (SI-52 4E)
- Separation of Phosphate, Phosphtite and Hypophosphite (SI-90 4E)
- Analysis of Sulfate According to Japanese Testing Methods for Fertilizers (SI-90 4E)
- Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (1)
- Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (2)
- Effects of Temperature on Retention Time (SI-35 4D)
- Analysis of Oxyhalides and Anions in Tap Water (SI-35 4D)
- Analysis of Organic Acids and Oxyhalides (SI-35 4D)
- Methanesulfonic Acid and Methanedisulfonic Acid (SI-35 4D)
- Calibration Curves (SI-35 4D)
- Effects of Injection Volume (SI-35 4D)
- Effects of Sample Concentration on Theoretical Plate Number (SI-35 4D)
- Analysis of Tricarboxylic Acids (SI-35 4D)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-35 4D)
- Analysis of Sulfite in Disposable Wooden Chopsticks (SI-35 4D)
- Analysis of Alkyl Sulfonic Acids (2) (SI-35 4D)
- Rapid Analysis of Anion Standards (SI-35 2B)
- Calibration Curves (SI-35 2B)
- Effects of Temperature on Retention Time (SI-35 2B)
- Analysis of Anions (SI-36 4D)
- Effects of Potassium Hydroxide Concentration on Retention Time (SI-36 4D)
- Effects of Potassium Hydroxide Concentration on Retention Time (2) (SI-36 4D)
- Effects of Temperature on Retention Time (SI-36 4D)
- Simultaneous Analysis of Common Oxyhalides and Anions (SI-37 4D)
- Analysis of Oxyhalides in Artificial-Drinking Water According to EPA Method 300.1 (SI-37 4D)