Analysis of Oxyhalides and Anions (SI-52 4E)

Please use this column with suppressor type ion chromatography system.

High resolution ion chromatography column, IC SI-52 4E was used for the quantification of several oxyhalides and anions. Ethylenediamine (0.1 % v/v of 50 mg/mL) is added to the water samples to prevent HCl production from the reaction between residual chloride and chlorous acid.



Sample : 50 μL
Anion std. (1 mg/L each) containing EDA 50 mg/L
1. F-
2. ClO2-
3. BrO3-
4. Cl-
5. HCO3-
6. NO2-
7. Br-
8. ClO3-
9. NO3-
10. PO43-
11. SO42-


Column       : Shodex IC SI-52 4E (4.0 mm I.D. x 250 mm)
Eluent       : 3.6 mM Na2CO3 aq.
Flow rate    : 0.8 mL/min
Detector     : Suppressed conductivity
Column temp. : 45 °C