Analysis of Tricarboxylic Acids (SI-35 4D)

Please use this column with suppressor type ion chromatography system.

Citric acids and aconitic acids which are tricarboxylic acids were analyzed using IC SI-35 4D (a column for anion analysis). In this analysis, citric acid and isocitric acid, cis-aconitic acid and trans-aconitic acid could be separated.


Sample : 20 μL
1. Citric acid 10 mg/L
2. Isocitric acid 50 mg/L
3. trans-Aconitic acid 20 mg/L
4. cis-Aconitic acid 20 mg/L


Column       : Shodex IC SI-35 4D (4.0 mm I.D. x 150 mm)
Eluent       : 9.0 mM Na2CO3 aq.
Flow rate    : 0.6 mL/min
Detector     : Suppressed conductivity
Column temp. : 45 °C