Calibration Curves (SI-35 4D)

Please use this column with suppressor type ion chromatography system.

The calibration curve for each ion, using IC SI-35 4D, an anion analysis column, has a coefficient of determination of ≥ 0.99 demonstrating high linearity in the range of 0.02 mg/L - 2 mg/L.


Sample : 100 μL
F-
Cl-
NO2-
Br-
NO3-
HPO42-
SO42-
ClO2-
ClO3-
BrO3-


Column       : Shodex IC SI-35 4D (4.0 mm I.D. x 150 mm)
Eluent       : 3.6 mM Na2CO3 aq.
Flow rate    : 0.6 mL/min
Detector     : Suppressed conductivity
Column temp. : 45 °C