Anion Standards and Organic Acids (3) (SI-50 4E)

Please use this column with suppressor type ion chromatography system.

Anion standards and organic acids were analyzed using IC SI-50 4E (a column for anions analysis).

Sample: 20 µL

  1. 1.F- 2 mg/L
  2. 2.Acetic acid 10 mg/L
  3. 3.Formic acid 2 mg/L
  4. 4.Methacrylic acid 10 mg/L
  5. 5.Cl- 3 mg/L
  6. 6.NO2- 5 mg/L
  7. 7.Br- 10 mg/L
  8. 8.NO3- 10 mg/L
  9. 9.HPO42- 15 mg/L
  10. 10.SO42- 15 mg/L
  11. 11.Oxalic acid 15 mg/L
Column
Shodex IC SI-50 4E (4.0 mm I.D. x 250 mm)
Eluent
3.2 mM Na2CO3 + 1.0 mM NaHCO3 aq
Flow rate
0.7 mL/min
Detector
Suppressed conductivity
Column temp.
25 ℃

Sample Name Index