Please use this column with suppressor type ion chromatography system.
The Japanese regulation for “Testing Methods for Fertilizers (2020*)” lists IC SI-90 4E as a suppressor type IC (anion) column for the analysis of sulfuric acid in fertilizer.
In this application, we used SI-90 4E to analyze sulfate in a standard solution and in a chemical fertilizer, one of the target fertilizer additives. The sample was prepared by following the extraction method mentioned in the “Research Report of Fertilizer Volume 13 (2020)”. A good linearity was obtained for the calibration curve in the concentration range 0.2 to 5 mg/mL.
*The version at the time of the application acquisition.
Sample preparation
1. Weigh 1.00-g of the target sample and put it in a 200-mL Erlenmeyer flask with stopper.
2. Add 100-mL hydrochloric acid and mix the contents for 10 minutes by placing the flask on a flask shaker.
3. Pour the supernatant into a 50-mL centrifuge tube with stopper, and centrifuge at 1,700 x g for 10 minutes.
4. Dilute the supernatant 1 + 99 with water. Then, further dilute the sample 1+19 with water.
5. Filter the mixture with a 0.45-µm membrane filter and use it as a sample.
Sample : 20 μL
Srandard 5 μg/mL
Extracted sample
1. SO42-
Column : Shodex IC SI-90 4E (4.0 mm I.D. x 250 mm) Eluent : 1.8 mM Na2CO3 + 1.7 mM NaHCO3 aq. Flow rate : 1.0 mL/min Detector : Suppressed conductivity Column temp. : 30 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Table of Anion Retention Time under Various Eluent Conditions (SI-90 4E)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-90 4E)
- Effects of Sodium Bicarbonate Concentration on Retention Time (SI-90 4E)
- Effects of Temperature on Elution Time (SI-90 4E)
- Effects of Added Acetonitrile on the Retention Time (SI-90 4E)
- Effects of Column Temperature on Retention Time (SI-50 4E)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-50 4E)
- Anion Standards (19) Rapid Analysis (SI-90 4E)
- Anion Standards (20) (Simultaneous Analysis of HS-, SO32-, SO42- and S2O32-) (SI-90 4E)
- Anion Standards (21) (Methanesulfonic Acid)
- High Sensitive Analysis of Anions (SI-90 4E)
- Anion Standards and Organic Acids (2) (SI-90 4E)
- Anion Standards and Organic Acids (3) (SI-50 4E)
- Mineral Water (2) (SI-90 4E)
- Sulfite in Wine (2) (SI-90 4E)
- Sodium Azide (2) (SI-90 4E)
- Sulfamic Acid (SI-90 4E)
- Sea Water (3) (SI-90 4E)
- Perchloric Acid (SI-90 4E)
- Analysis of Iodide Ion (SI-90 4E)
- Fluoroacetid Acids (SI-50 4E)
- Anion Analysis in Lithium Salts (3) (SI-90 4E)
- Analysis of Oxyhalides (SI-52 4E)
- Effects of Temperature on Retention Time (SI-52 4E)
- Calibration Curves for Oxyhalides (SI-52 4E)
- Analysis of Organic Acids and Oxyhalides (SI-52 4E)
- Analysis of Oxyhalides and Anions (SI-52 4E)
- Analysis of Oxyhalides and Anions in River Water (SI-52 4E)
- Analysis of Oxyhalides and Anions in Rain Water (SI-52 4E)
- Analysis of Oxyhalides and Anions in Tap Water (SI-52 4E)
- Iodic Acid (SI-52 4E)
- Separation of Phosphate, Phosphtite and Hypophosphite (SI-52 4E)
- Separation of Phosphate, Phosphtite and Hypophosphite (SI-90 4E)
- Analysis of Chloride According to Japanese Testing Methods for Fertilizers (SI-52 4E)
- Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (1)
- Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (2)
- Effects of Temperature on Retention Time (SI-35 4D)
- Analysis of Oxyhalides and Anions in Tap Water (SI-35 4D)
- Analysis of Organic Acids and Oxyhalides (SI-35 4D)
- Methanesulfonic Acid and Methanedisulfonic Acid (SI-35 4D)
- Calibration Curves (SI-35 4D)
- Effects of Injection Volume (SI-35 4D)
- Effects of Sample Concentration on Theoretical Plate Number (SI-35 4D)
- Analysis of Tricarboxylic Acids (SI-35 4D)
- Effects of Sodium Carbonate Concentration on Retention Time (SI-35 4D)
- Analysis of Sulfite in Disposable Wooden Chopsticks (SI-35 4D)
- Analysis of Alkyl Sulfonic Acids (2) (SI-35 4D)
- Rapid Analysis of Anion Standards (SI-35 2B)
- Calibration Curves (SI-35 2B)
- Effects of Temperature on Retention Time (SI-35 2B)
- Analysis of Anions (SI-36 4D)
- Effects of Potassium Hydroxide Concentration on Retention Time (SI-36 4D)
- Effects of Potassium Hydroxide Concentration on Retention Time (2) (SI-36 4D)
- Effects of Temperature on Retention Time (SI-36 4D)
- Simultaneous Analysis of Common Oxyhalides and Anions (SI-37 4D)
- Analysis of Oxyhalides in Artificial-Drinking Water According to EPA Method 300.1 (SI-37 4D)