Anion Standards (19) Rapid Analysis (SI-90 4E)

Please use this column with suppressor type ion chromatography system.

We used an IC SI-90 4E, a column for anion analysis, to analyze seven standard anions (F-, Cl-, NO2-, Br-, NO3-, HPO42-, SO42-). The use of an eluent with aqueous solution of 1.8 mM Na2CO3 + 1.7 mM NaHCO3 is best suited to the separation of those seven anions. A good separation of seven ions and carbonate ion was achieved even at 2.0 mL/min flow rate, demonstrating the method’s suitability for a rapid analysis.

Sample: 20 μL
Standard 1 mg/L (in DMSO)

  1. 1.F- (2 mg/L)
  2. 2.Cl- (3 mg/L)
  3. 3.NO2- (5 mg/L)
  4. 4.Br- (10 mg/L)
  5. 5.NO3- (10 mg/L)
  6. 6.HCO3- (300 mg/L)
  7. 7.HPO42- (15 mg/L)
  8. 8.SO42- (15 mg/L)
Column
Shodex IC SI-90 4E (4.0 mm I.D. x 250 mm)
Eluent
1.8 mM Na2CO3 + 1.7 mM NaHCO3 aq.
Flow rate
1.0 mL/min
Detector
Suppressed conductivity
Column temp.
30 ℃