Analysis of Organic Acids and Oxyhalides (SI-52 4E)

Please use this column with suppressor type ion chromatography system.

High resolution ion chromatography column, IC SI-52 4E was used for the quantification of several oxyhalides and organic acids. Marlic acid and succinic acid will coelute with malonic acid under this condition.

 

 

Sample : 50 μL
1. F- 1 mg/L
2. Acetic acid 0.5 mg/L
3. Propionic acid 0.8 mg/L
4. Formic acid 0.1 mg/L
5. ClO2- 1 mg/L
6. BrO3- 1 mg/L
7. Cl- 1 mg/L
8. NO2- 1 mg/L
9. Br- 1 mg/L
10. ClO3- 1 mg/L
11. NO3- 1 mg/L
12. HPO42- 1 mg/L
13. SO42- 1 mg/L
14. Malonic acid 0.5 mg/L


 

Column       : Shodex IC SI-52 4E (4.0 mm I.D. x 250 mm)
Eluent       : 3.6 mM Na2CO3 aq.
Flow rate    : 0.8 mL/min
Detector     : Suppressed conductivity
Column temp. : 45 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index