Analysis of Organic Acids and Oxyhalides (SI-52 4E)

Please use this column with suppressor type ion chromatography system.

High resolution ion chromatography column, IC SI-52 4E was used for the quantification of several oxyhalides and organic acids. Marlic acid and succinic acid will coelute with malonic acid under this condition.

Sample: 50 µL

  1. 1.F- 1 mg/L
  2. 2.Acetic acid 0.5 mg/L
  3. 3.Propionic acid 0.8 mg/L
  4. 4.Formic acid 0.1 mg/L
  5. 5.ClO2- 1 mg/L
  6. 6.BrO3- 1 mg/L
  7. 7.Cl- 1 mg/L
  8. 8.NO2- 1 mg/L
  9. 9.Br- 1 mg/L
  10. 10.ClO3- 1 mg/L
  11. 11.NO3- 1 mg/L
  12. 12.HPO42- 1 mg/L
  13. 13.SO42- 1 mg/L
  14. 14.Malonic acid 0.5 mg/L
Column
Shodex IC SI-52 4E (4.0 mm I.D. x 250 mm)
Eluent
3.6 mM Na2CO3 aq.
Flow rate
0.8 mL/min
Detector
Suppressed conductivity
Column temp.
45 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications