Calibration Curves of Hydrophilic Polymers (2) (Sodium Polyacrylate)

Calibration curves for Asahipak GF-HQ series using sodium polyacrylate are shown. Since Asahipak GF-HQ series can be used with both aqueous and organic solvents as the eluent, it is suitable for the analysis of hydrophilic polymers such as sodium poly(styrene sulfonate) or sodium polyacrylate. Several types of hydrophilic polymer standard are commercially available. It is recommended to use standard polymer similar to the analyte for the calibration curve.


Sample :
Sodium polyacrylate


Columns      : Shodex Asahipak GF-HQ series (7.5 mm I.D. x 300 mm each)
Eluent       : 50 mM Na2HPO4 aq./CH3CN=90/10
Flow rate    : 0.6 mL/min
Detector     : RI
Column temp. : 40 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)