Selection of Column (3) (Poly(Sodium Methacrylate))

Calibration curves of Asahipak GF-7M HQ (packed with poly(vinyl alcohol)) and OHpak SB-806M HQ (packed with poly(hydroxy methacrylate)) are compared using a hydrophilic polymer standerd, poly(sodium methacrylate). The comparison shows that the calibration curve of Asahipak GF-7M HQ is more linear than that of OHpak SB-806M HQ.


Sample : Poly(sodium methacrylate) standard


Columns      : Shodex Asahipak GF-7M HQ (7.5 mm I.D. x 300 mm),
               OHpak SB-806M HQ (8.0 mm I.D. x 300 mm)
Eluent       : 50 mM Na2HPO4 aq./CH3CN=90/10
Flow rate    : 0.6 mL/min
Detector     : RI
Column temp. : 40 °C