Analysis of Pyridylaminated Sugar Chains (NH2P-50 4E)

Pyridylaminated (PA) sugar chains were separated using Asahipak NH2P-50 4E. Refractive Index detector (RI) is generally used for sugar analyses, however pyridylaminated sugar chains can be detected by more sensitive fluorescence detectors. Also the analysis can be carried under gradient elution.


Sample : 0.125 pmol/μL, 5 μL
1. PA-Sugar chain 012
2. PA-Sugar chain 013
3. PA-Sugar chain 014
4. PA-Sugar chain 001
5. PA-Sugar chain 015
6. PA-Sugar chain 002
7. PA-Sugar chain 010
8. PA-Sugar chain 004

Column       : Shodex Asahipak NH2P-50 4E (4.6 mm I.D. x 250 mm)
Eluent       : (A); 200 mM Acetic acid-Triethylamine (pH7.3)/CH3CN=35/65
               (B); 200 mM Acetic acid-Triethylamine (pH7.3)/CH3CN=50/50
               Linear gradient: 0 to 60 min,  A to B
Flow rate    : 0.5 mL/min
Detector     : Fluorescence (Ex: 310 nm , Em: 380 nm)
Column temp. : 40 °C