A polymer based amino column, Asahipak NH2P-40 2D, was used for LC/MS/MS analysis of saccharides and sugar alcohols. Adding ammonia to the eluent enhances the ionization of saccharides, thus providing high sensitivity detection.
*Sales of Asahipak NH2P-40 2D used in this application is discontinued. Please consider trying Asahipak NH2P-50 2D.
Sample: 5 μL
50 ng/mL each
- 1.meso-Erythritol
- 2.Xylitol
- 3.Fructose
- 4.Glucose
- 5.Sucrose
- 6.Lactose
- 7.Maltose
- 8.Raffinose

- Column
- :Shodex Asahipak NH2P-40 2D (2.0 mm I.D. x 150 mm)
- Eluent
- :(A) 0.05 % NH3 aq./(B) CH3CN
Isocratic; (B %) 75 % - Flow rate
- :0.2 mL/min
- Detector
- :ESI-MS/MS (MRM Negative)
(m/z)
121→82(-) for meso-Erythritol, 151→59(-) for Xylitol, 179→89(-) for Fructose and Glucose,
341→59(-) for Sucrose, 341→161(-) for Lactose and Maltose, 503→89(-) for Raffinose - Column temp.
- :40 ℃
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Analysis of Pyridylaminated (PA) Monosaccharides (NH2P-50 4E)
- Analysis of Pyridylaminated Sugar Chains (NH2P-50 4E)
- LC/TOF-MS Analysis of 2-Aminobenzoic Acid Labeled N-Glycan (Human IgG) (NH2P-40 2D)
- Analysis of 2-Aminobenzoic Acid Labeled Glucose Homopolymer (NH2P-50 4E)
- LC/MS/MS Analysis of Saccharides (VG-50 2D)
- LC/MS Analysis of Phosphorylated Saccharides (VT-50 2D)
- Simultaneous Analysis of Saccharides, Organic Acids and Amino Acids with LC/MS (VG-50 2D)
- Simultaneous Analysis of Hydrophilic Substances in Energy Drink with LC/MS (VG-50 2D)
- LC/MS Analysis of Glucose and Gluconic Acid (VG-50 2D)
- LC/MS Analysis of N-Acetylchitooligosaccharides (NH2P-40 2D)