Analysis of Contaminants Sucrose and Sorbitol in Cranberry Juice According to USP-NF Method (SC1011-7F)

According to the United States Pharmacopeia and the National Formulary (USP-NF 2022 ISSUE 3* Effective December 1, 2022), analysis of contaminants sucrose and sorbitol in cranberry juice should be carried out with a column with L19 packing material and meets following requirements. The EP SC1011-7F confirmed the requirements were met.

 

System suitability requirements

Resolution of sucrose and sorbitol: ≥ 1.8
Relative standard deviation (RSD): ≤ 2.0 %

*The version at the time of the application acquisition.

Sample: 20 μL
0.1 mg/mL each

  1. 1.Sucrose
  2. 2.Sorbitol
Column
Shodex EP SC1011-7F (7.8 mm I.D. x 300 mm)
Eluent
H2O
Flow rate
0.5 mL/min
Detector
RI
Column temp.
85 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications