Analysis of Azithromycin for Injection According to USP-NF Method (ODP-50 4D)

Azithromycin is a 15-membered ring macrolide antibiotic. According to the United States Pharmacopeia and the National Formulary (USP 42-NF 37*), "ASSAY" for "Azithromycin for Injection" should be carried out with a analysis column and its guard column filled with L67 packing material. The Asahipak ODP-50 4D and ODP-50G 4A (guard column) filled with L76 packing material, in series confirmed the requirements were met.

System suitability requirementsResolution between azithromycin A and azaerythromycin (System suitability solution): ≥ 2.5
Tailing factor (Standard solution): 0.9 - 1.5
Relative standard deviation (RSD) (Standard solution): ≤ 2%

*The version at the time of the application acquisition.

Sample: 15 µL
(Standard solution) 1 mg/mL of Azaerythromycin in H2O/CH3CN=48/52
(System suitability solution) 1 mg/mL each of Azaerythromycin A and Azithromycin in H2O/CH3CN=48/52

  1. 1.Azaerythromycin A
  2. 2.Azaerythromycin
Column
Shodex Asahipak ODP-50G 4A (4.6 mm I.D. x 10 mm) + ODP-50 4D (4.6 mm I.D. x 150 mm)
Eluent
6.7 g/L Dibasic potassium phosphate aq./CH3CN=48/52 (adjusted to pH11.0 with 10 M KOH aq.)
Flow rate
1.0 mL/min
Detector
UV (215 nm)
Column temp.
40 ℃

Sample Name Index