The figure shows calibration curves of IC NI-424 (an anion analysis column for non-suppressor method) for three kinds of anions. The calibration curves for chloride and nitrite ions exhibit linearity in a range between 0.01 and 100 mg/L. For phosphoric acid ion which has low sensitivity, the calibration curve exhibits linearity in a range between 0.1 and 100 mg/L.
Sample : 50 µL
1. H2PO4-
2. Cl-
3. NO2-
Column : Shodex IC NI-424 (4.6 mm I.D. x 100 mm) Eluent : 2.1 mM Phthalic acid + 2.9 mM Amino-n-capronic acid + 6 mM Phenylboronic acid aq. Flow rate : 1.0 mL/min Detector : Non-suppressed conductivity Column temp. : 40 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Effects of Sample Load on Column Efficiency (4) (NI-424)
- Calibration Curves for Low Concentration F- (NI-424)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Anion Standards (16) (NI-424: Phthalic Acid aq. Eluent)
- Anion Standards (17) (Hydrophobic Inorganic Ions) (NI-424)
- Anion Standards (18) (NI-424)
- Tap Water (7) (NI-424)
- River Water (3) (NI-424)
- Rain Water (NI-424)
- Phosphonates (NI-424)
- Effects of pH on Retention Time (NI-424)
- Analysis of Pyrophosphate and Tripolyphosphate (NI-424)
- Carbonate and Other Anions (2) (NI-424)
- Anions and Organic Acids (NI-424)
- Methanesulfonic Acid and Methanedisulfonic Acid (NI-424)
- Elution Volume of Organic Acids (NI-424)
- Analysis of Alkyl Sulfonic Acids (1) (NI-424)
- Anion Analysis in Lithium Salts (1) (NI-424)
- Analysis of Sulfamic Acid and Ammonium Thiocyanate According to Japanese Testing Methods for Fertilizers (NI-424)
- Analysis of Chloride According to Japanese Testing Methods for Fertilizers (NI-424)
- Sulfate Ion in Enoxaparin Sodium (NI-424)