Anion Standards (15) (Comparison of NI-424 and I-524A)

The seven main anions can be separated using an anion analysis column for non-suppressor method IC NI-424. The theoretical plate number of the column is more than double that of the existing column IC I-524A.
Comparison with I-524A column:
1) NI-424 column has achieved separation of phosphoric acid ion and fluoride ion which was difficult to sepatate using I-524A.
2) With NI-424, each peak is sharper, and separation of ions and peak balance have been improved. In particular, improvement has been made in the separation of chloride and nitrite ions making NI-424 column suited to the analysis of a trace amount of nitrite ion contained in high-concentration chloride ion; demand for such analysis is high in the area of environmental analysis.


Sample : 20 μL
1. H2PO4- 10 mg/L
2. F- 1 mg/L
3. Cl- 1 mg/L
4. NO2- 5 mg/L
5. Br- 5 mg/L
6. NO3- 5 mg/L
7. SO42- 5 mg/L


 

(Left)
Column       : Shodex IC I-524A (4.6 mm I.D. x 100 mm)
Eluent       : 2.5 mM Phthalic acid + 2.3 mM Tris(hydroxymethyl)aminomethane aq.
Flow rate    : 1.2 mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40 °C
(Right)
Column       : Shodex IC NI-424 (4.6 mm I.D. x 100 mm)
Eluent       : 8 mM 4-Hydroxybenzoic acid + 2.8 mM Bis-Tris + 2 mM Phenylboronic acid
               + 0.005 mM *CyDTA aq.
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40 °C
* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid