Tap Water (7) (NI-424)

The administrative order of the Ministry of Health, Labour and Welfare #261 (July 22, 2003) specifies the acceptable levels for F-, Cl-, NO2- and NO3-. Their concentrations are measured using ion chromatography. Here, anions in tap water were analyzed using IC NI-424 (an anion analysis column for non-suppressor method) though NO2- could not be detected in tap water.
Please refer to 'Regulation for Water Supply '.


Sample : Tap water, 100 μL
1. F-
2. Cl-
3. NO3-
4. SO42-


Column       : Shodex IC NI-424 (4.6 mm I.D. x 100 mm)
Eluent       : 8 mM 4-Hydroxybenzoic acid + 2.8 mM Bis-Tris + 2 mM Phenylboronic acid 
               + 0.005 mM *CyDTA aq.
 
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40 °C
* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid