Effects of pH on Retention Time (NI-424)

Effects of eluent pH on the elution of each anion using IC NI-424 (an anion analysis column for non-suppressor method) are shown. When the pH of the eluent is increased, the retention time for the anions is decreased.


Sample : 20 μL
H2PO4- 10 mg/L
F- 10 mg/L
Cl- 10 mg/L
NO2- 10 mg/L
Br- 10 mg/L
NO3- 10 mg/L
I- 20 mg/L
SO42- 10 mg/L
S2O32- 50 mg/L
SCN- 50 mg/L
ClO4- 50 mg/L


Column       : Shodex IC NI-424 (4.6 mm I.D. x 100 mm)
Eluent       : 8 mM 4-Hydroxybenzoic acid + 2 mM Phenylboronic acid
               + 0.005 mM *CyDTA (pH adjusted with Bis-Tris)
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40 °C

*CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid