IC NI-424 (an anion analysis column for non-suppressor method) has high theoretical plate number (more than 5,000). Carbonate and other anions can be separated well with NI-424.
Sample : 20 μL
1. Acetic acid 20.0 mg/L
2. Formic acid 20.0 mg/L
3. Malic acid 20.0 mg/L
4. Tartaric acid 20.0 mg/L
5. Oxalic acid 20.0 mg/L
6. H2PO4- 20.0 mg/L
7. F- 1.5 mg/L
8. Cl- 2.5 mg/L
9. NO2- 6.0 mg/L
10. Br- 10.0 mg/L
11. NO3- 10.0 mg/L
12. SO42- 10.0 mg/L
Column : Shodex IC NI-424 (4.6 mm I.D. x 100 mm) Eluent : 8 mM 4-Hydroxybenzoic acid + 2.8 mM Bis-Tris + 2 mM Phenylboronic acid + 0.005 mM *CyDTA aq. Flow rate : 1.0 mL/min Detector : Non-suppressed conductivity Column temp. : 40 °C
* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Effects of Sample Load on Column Efficiency (4) (NI-424)
- Linearity of Calibration Curves (NI-424)
- Calibration Curves for Low Concentration F- (NI-424)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Anion Standards (16) (NI-424: Phthalic Acid aq. Eluent)
- Anion Standards (17) (Hydrophobic Inorganic Ions) (NI-424)
- Anion Standards (18) (NI-424)
- Tap Water (7) (NI-424)
- River Water (3) (NI-424)
- Rain Water (NI-424)
- Phosphonates (NI-424)
- Effects of pH on Retention Time (NI-424)
- Analysis of Pyrophosphate and Tripolyphosphate (NI-424)
- Carbonate and Other Anions (2) (NI-424)
- Methanesulfonic Acid and Methanedisulfonic Acid (NI-424)
- Elution Volume of Organic Acids (NI-424)
- Analysis of Alkyl Sulfonic Acids (1) (NI-424)
- Anion Analysis in Lithium Salts (1) (NI-424)
- Analysis of Sulfamic Acid and Ammonium Thiocyanate According to Japanese Testing Methods for Fertilizers (NI-424)
- Analysis of Chloride According to Japanese Testing Methods for Fertilizers (NI-424)
- Sulfate Ion in Enoxaparin Sodium (NI-424)