Biological sample was analyzed directly using ODP2 HP. ODP2 HP is polymer based reversed phase column. One of advantage of ODP2 HP is very low adsorption of protein, because the surface of the packing material is high polar and the pore size is small. Since proteins like BSA are eluted early and its peak shape is sharp. ODP2 HP can separate drugs from proteins. Four LC/MS chromatograms of barbital were shown. These are comparison of using ODP2 HP and ODS, and barbital without BSA and with BSA. In using ODS (right), the peak height of barbital with BSA were lower than without BSA. It means there was ion suppression by existence of BSA. The comparison data of recovery of barbital also showed influence of BSA to ODS. ODP2 HP can be used for analysis of low molecular compounds in biological sample without bad influence of protein.
Sample : Barbital (500 ng/L) in BSA (7.0 mg/L), 10 μL
1. Barbital
Sample : Barbital (500 ng/L) in BSA (7.0 mg/L), 10 μL
1. Barbital
Column : Shodex ODP2 HP-2B, ODS from other manufacturer(2.0 mm I.D. x 50 mm each) Eluent : 10 mM Ammonium acetate/CH3CN=70/30 Flow rate : 0.2 mL/min Detector : ESI-MS (SIM negative) Column temp. : 30 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Indomethacin in Serum (ODP-50 6E)
- Procainamide in Serum (ODP-50 4D)
- LC/MS Analysis of Drugs in BSA (1) (ODP2 HP)
- LC/MS Analysis of Drugs in BSA (2) (ODP2 HP)
- LC/MS/MS Analysis of Tricyclic Antidepressants in BSA (ODP2 HP)
- LC/MS Analysis of Drugs in Control Serum (ODP2 HP)
- Reproducibility in LC/MS Analysis (ODP2 HP)
- Influence of Proteins in LC/MS Analysis (ODP2 HP-2B)
- Direct Analysis of Acetaminophen in Serum (ODP2 HP)
- Direct Analysis of Risperidone in Urine (ODP2 HP)
- Simultaneous Analysis of Multi Symptom Cold Formula in Serum (ODP2 HP)
- Analysis of Anticonvulsant in Serum (ODP2 HP)
- Hippuric Acid etc. in Urine (2)