LC/MS analysis of drugs in control serum after removing proteins was done by polymer-based column ODP2 HP. Each drug has peak areas between 82 - 111 % (peak areas without control serum are set as 100 %), which means a good separation without receiving ion suppression from protein.
Sample : drugs 500 ng/mL each in Control serum, 5 μL
Lidocaine
Chlorpheniramine
Propranolol
Diphenhydramine
Bupivacaine
Verapamil
Imipramine
Column : Shodex ODP2 HP-2B (2.0 mm I.D. x 50 mm) Eluent : 10 mM Ammonium acetate aq./CH3CN=70/30 Flow rate : 0.2 mL/min Detector : UV (280 nm), ESI-MS (SIM Positive) Column temp. : 40 °C