Various steviol glycosides which are used as sweetener were analyzed using Asahipak NH2P-40 2D, a polymer-based amino column, with LC/MS/MS detection. Steviolbioside and Rebaudioside B which have carboxyl group can be eluted under alkaline conditions. Furthermore, alkaline conditions promote the deprotonation of hydroxyl groups at the time of ionization. Therefore, alkaline conditions are suitable for high sensitive detection of substances with hydroxyl groups such as saccharides under the negative mode.
*Sales of the product used in this application is discontinued, however if you are still interested in this application, please contact us.
Sample : 5 μL
50 ng/mL each
(in H2O/CH3CN=1/1)
Isosteviol
Stevioside
Rebaudioside C
Rebaudioside F
Rebaudioside A
Steviolbioside
Rebaudioside D
Rebaudioside B
Column : Shodex Asahipak NH2P-40 2D (2.0 mm I.D. x 150 mm) Eluent : (A); 0.1 % NH3 aq./(B); CH3CN Isocratic ; (B %) 78 % Flow rate : 0.2 mL/min Detector : ESI-MS (SIM Negative) Column temp. : 30 °C
Sample Name Index
Applications
- Acesulfame K (SC1011)
- Analysis of Acesulfame K and Sucralose (SC1011)
- Sucralose (SP0810)
- Stevioside and Rebaudioside A (NH2P-50 4E)
- Quantification of Purified Stevia Extract (VG-50 4E)
- Steviol Purity Test of Purified Stevia Extracts According to Japanese Pharmaceutical Excipients Method (C18M 4E)
- Analysis of Mogroside V in Momordica Grosvenori (NH2P-50 4E)
- LC/MS Analysis of Various Sweeteners (NH2P-40 2D)
- Simultaneous Analysis of Sweeteners and Amino Acids in Energy Drink (NH2P-40 2D)
- Simultaneous Analysis of Saccharides and Sweeteners in Yogurt Drink (NH2P-40 2D)
- Effects of Eluent pH on HILIC Mode (VG-50 2D)
- Analysis of Thaumatin (SP-825)