Sweeteners in an energy drink were analyzed using Asahipak NH2P-40 2D, a polymer-based amino column, with LC/MS/MS detection. Under these conditions, sweeteners, amino acids and water-soluble vitamins such as carnitine and pyridoxine can be simultaneously analyzed.
*Sales of Asahipak NH2P-40 2D used in this application is discontinued. Please consider trying Asahipak NH2P-50 2D.
Sample: 5 μL
500-fold diluted Energy drink with H2O/CH3CN=1/1
- Sucralose
- Acesulfame (k)
- meso-Erythritol
- Carnitine
- Leucine
- Isoleucine
- Valine
- Pyridoxine
- Rebaudioside A

- Column
- :Shodex Asahipak NH2P-40 2D (2.0 mm I.D. x 150 mm)
- Eluent
- :(A) 0.1 % NH3 aq./(B) CH3CN
Isocratic; (B %) 78 % - Flow rate
- :0.2 mL/min
- Detector
- :ESI-MS (SIM Positive/Negative)
- Column temp.
- :30 ℃
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Acesulfame K (SC1011)
- Analysis of Acesulfame K and Sucralose (SC1011)
- Sucralose (SP0810)
- Stevioside and Rebaudioside A (NH2P-50 4E)
- LC/MS Analysis of Various Steviol Glycosides (NH2P-40 2D)
- Quantification of Purified Stevia Extract (VG-50 4E)
- Steviol Purity Test of Purified Stevia Extracts According to Japanese Pharmaceutical Excipients Method (C18M 4E)
- Analysis of Mogroside V in Momordica Grosvenori (NH2P-50 4E)
- LC/MS Analysis of Various Sweeteners (NH2P-40 2D)
- Simultaneous Analysis of Saccharides and Sweeteners in Yogurt Drink (NH2P-40 2D)
- Effects of Eluent pH on HILIC Mode (VG-50 2D)
- Analysis of Thaumatin (SP-825)