Asahipak GS-320 HQ (a multimode column) is suitable not only for the separation of nucleobases, nucleosides and nucleotides but also for the separation of samples of their mixtures. GS-320 HQ permits one-time separation by isocratic elution without using gradient elution.
Sample : 20 μL
1. ATP 100 μg/mL
2. ADP 100 μg/mL
3. IMP 100 μg/mL
4. AMP 50 μg/mL
5. GMP 100 μg/mL
6. Inosine 100 μg/mL
7. Adenine 100 μg/mL
8. Hypoxanthine 100 μg/mL
9. Guanine 100 μg/mL
10. Adenosine 100 μg/mL
Column : Shodex Asahipak GS-320 HQ (7.5 mm I.D. x 300 mm) Eluent : 205 mM NaH2PO4 aq./205 mM H3PO4 aq.=300/7 Flow rate : 0.6 mL/min Detector : UV (260 nm) Column temp. : 30 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Nucleobases and Nucleosides (1) (DE-613)
- Nucleobases and Nucleosides (2) (NN-814)
- Nucleobases (1) (GS-320 HQ)
- Nucleobases (4) (DE-413)
- Guanine and Tryptophan (DM-614, NN-814 and DE-613)
- Adenosine and Derivatives
- "Umami" (GS-320 HQ)
- Nucleosides (1) (GS-320 HQ)
- Adenine Nucleotides in Red Blood Cell
- Purine Bases and Uric Acid Separation (GS-320 HQ)