Monovalent and divalent cations were analyzed using IC YS-50 . 18-Crown 6- Ether will make a complex with cations, whose complex building depends on the kind of cations. Therefore, adding 18-Crown-6-Ether is able to control the elusion of cations and improve the separation of Na+ and NH4+.


Sample : 10 μL
1. Li+ 2 mg/L
2. Na+ 10 mg/L
3. NH4+ 10 mg/L
4. K+ 20 mg/L
5. Mg2+ 10 mg/L
6. Ca2+ 20 mg/L

 

Column       : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Eluent       : 4 mM Methanesulfonic acid + n mM 18-Crown 6-ether aq.
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40 °C