Effects of Flow Rate on Plate Number (YS-50)

Although the optimal flow rate is 0.8 or less mL/min, there are few falls of theoretical plate number at 1.0 mL/min, and since ca2+ can be analyzed within 20 minutes, by general cation analysis using IC YS-50 , We recommend to set flow rate below to 1.0 mL/min.


Sample : 10 μL
1. Li+ 2 ppm
2. Na+ 10 ppm
3. NH4+ 10 ppm
4. K+ 20 ppm
5. Mg2+ 10 ppm
6. Ca2+ 20 ppm


Column       : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Eluent       : 4 mM Methanesulfonic acid aq. 
Detector     : Non-suppressed conductivity
Column temp. : 40 °C