Monovalent and Divalent Cations Analysis in Tap Water (Ethylenediamine Containing Sample) (YS-50)

Chloric acid should be analyzed by ion chromatography for water quality standard in accordance to the Ministry of Health, Labor and Welfare in Japan. 1 mL of Ethylendiamine (EDA) solution (50 mg/mL) is to be added into 1 L sample, to prevent generation of chloric acid through the reaction between chlorous acid and residual chlorine. Monovalent and divalent cations in tap water were analyzed by IC YS-50 .



Sample : 50 μL
Cation std. containing EDA 50 mg/L
1. Li+ 2 mg/L
2. Na+ 10 mg/L
3. NH4+ 10 mg/L
4. K+ 20 mg/L
5. Mg2+ 10 mg/L
6. Ca2+ 20 mg/L
7. Ethylenediamine, EDA 50 mg/L

Sample : 50 μL
Tap water containing EDA 50 mg/L
2. Na+
3. NH4+
4. K+
5. Mg2+
6. Ca2+
7. Ethylenediamine, EDA 50 mg/L


Column       : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Eluent       : 4 mM Nitric acid + 1.5 mM 18-Crown 6-ether aq./CH3CN=90/10
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40 °C