Monovalent and divalent cation standards were analyzed using IC YK-421. This data shows the change in dynamic range of the column with the height of the peaks. There is the linearity in the 1 to 200 mg/L range for Na+ and K+ and in the 1 to 100 mg/L range for Ca2+ and Mg2+.


Sample :
Ca2+
K+
Mg2+
Na+


Column       : Shodex IC YK-421 (4.6 mm I.D. x 125 mm)
Eluent       : 5 mM Tartaric acid + 1 mM Dipicolinic acid + 1.5 g/L Boric acid aq.
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)