Analysis of Tetraalkylammoniums (YS-50)

Trialkylammoniums were separated using a cation chromatography column, IC YS-50 with charged aerosol detector (CAD). The column demonstrates low 'bleed' providing a stable baseline. The use of a low concentration of acetonitrile in the eluent will allow the tetraalkylammonium compounds to elute by hydrophobic interaction.

Sample: 5 µL

  1. 1.Tetramethylammonium 25 µg/mL
  2. 2.Tetraethylammonium 50 µg/mL
  3. 3.Tetrapropylammonium 100 µg/mL
  4. 4.Tetrabutylammonium 100 µg/mL
  5. 5.Tetrapentylammonium 100 µg/mL
  6. 6.Tetrahexylammonium 200 µg/mL
Column
Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Eluent
4 mM Nitric acid aq./CH3CN=65/35
Flow rate
1.0 mL/min
Detector
Charged aerosol
Column temp.
40 ℃