Red wine diluted 20 times was analyzed using IC YS-50 (a column for cation chromatography) by non-supprresor method. Monoethanolamine other than Na+, NH4+, K+, Mg2+ and Ca2+ was detected.


Sample : 20 fold diluted Red wine ,
10 μL
1. Na+
2. NH4+
3. Monoethanolamine
4. K+
5. Mg2+
6. Ca2+


Column       : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Eluent       : 4 mM Methanesulfonic acid aq.
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40 °C