Scroll
HK-400 |
- Newly developed styrene divinylbenzene copolymer monodisperse particles
- Analysis time is reduced to about a sixth of conventional column's analysis time
- Low column pressure even under high flow rate does not require its use with UHPLC device
- The amount of solvent used is reduced to about a sixth
- Fulfills USP L21 requirements
|
(High performance semi-micro columns) [Housing material: SUS]
HK-400 series is recommended to be used with semi-micro type devices.Scroll
Product code |
Product name |
Plate number
(TPN/column) |
Particle size (μm) |
Column size (mm)
I.D. x Length |
F6025010 |
GPC HK-401 |
≥ 9,000 |
3 |
4.6 x 150 |
F6025020 |
GPC HK-402 |
≥ 12,000 |
3 |
4.6 x 150 |
F6025030 |
GPC HK-403 |
≥ 9,000 |
3.5 |
4.6 x 150 |
F6026040 |
GPC HK-404L |
≥ 9,000 |
3.5 |
4.6 x 150 |
F6025050 |
GPC HK-405 |
≥ 7,000 |
3 |
4.6 x 150 |
F6025060 |
GPC HK-406 |
≥ 5,000 |
6.5 |
4.6 x 150 |
- HK-404L is a mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution.
Scroll
Product name |
Maximum pressure
(MPa/Column) |
Flow rate (mL/min) |
Temperature (℃) |
Shipping solvent |
Recommended |
Maximum |
Recommended |
Maximum |
GPC HK-401 |
25 |
0.3 - 1.0 |
2.0 |
40 |
60 |
Tetrahydrofuran (THF) |
GPC HK-402 |
20 |
0.3 - 1.0 |
1.5 |
40 |
60 |
Tetrahydrofuran (THF) |
GPC HK-403 |
25 |
0.3 - 1.0 |
2.0 |
40 |
60 |
Tetrahydrofuran (THF) |
GPC HK-404L |
25 |
0.3 - 1.0 |
2.0 |
40 |
60 |
Tetrahydrofuran (THF) |
GPC HK-405 |
25 |
0.3 - 1.0 |
2.0 |
40 |
60 |
Tetrahydrofuran (THF) |
GPC HK-406 |
25 |
0.3 - 1.0 |
2.0 |
40 |
60 |
Tetrahydrofuran (THF) |
(Guard filter for GPC HK series)Scroll
Product code |
Product name |
Contents |
F6700200 |
GPC HK-G |
One holder and one filter |
F6700100 |
GPC HK-G filter |
3 filters |
-
- - Removes sample-origin insoluble components.
- Attach directly to the analytical column.
Usable solventsScroll
Solvents |
Product name |
HK-401 |
HK-402 |
HK-403 |
HK-404L |
HK-405 |
HK-406 |
Chloroform |
✓ |
✓ |
✓ |
✓ |
✓ |
✓ |
N,N-Dimethylformamide (DMF) |
✓ |
✓ |
✓ |
✓ |
✓ |
✓ |
Toluene |
✓ |
✓ |
✓ |
✓ |
✓ |
✓ |
Hexafluoroisopropanol (HFIP) |
✓ |
× |
✓ |
✓ |
✓ |
✓ |
30 % HFIP/Chloroform |
✓ |
✓ |
✓ |
✓ |
✓ |
✓ |
- ✓: Can be used ×: Cannot be used
Target molecular weight range and exclusion limitPlease use the following tables for reference purposes only when selecting columns. Scroll
Product name |
Measured with polystyrene (eluent: THF) |
Target molecular weight range |
Exclusion limit |
GPC HK-401 |
100 - 1,500 |
2,000 |
GPC HK-402 |
200 - 10,000 |
20,000 |
GPC HK-403 |
2,000 - 70,000 |
100,000 |
GPC HK-404L |
100 - 1,000,000 |
1,000,000 |
GPC HK-405 |
10,000 - 2,500,000 |
4,000,000 |
GPC HK-406 |
30,000 - 8,000,000 |
10,000,000 |
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Related Products
Product Information (Related Information)