To exchange/replace the solvent currently filled in the Shodex GPC series columns, follow the below solvent exchange procedures.
Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume.
- Check the solvent replaceability of GPC series columns.
Refer to Solvent Replacement Applicability of Organic SEC (GPC) Columns. - Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column.
Refer to Solubility of Solvents.
- (1)When replacing the current solvent with a miscible solvent, first, introduce 1:1 mixture of the current solvent and the new solvent, and then replace it with 100 % of the new solvent.
e.g. When replacing THF with Chloroform
First introduce 1:1 mixture of THF and Chloroform, and then introduce 100 % Chloroform. - (2)When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent.
e.g. When replacing THF with “HFIP with CF3COONa”
First follow the example in procedure (1), and then replace 100 % HFIP with “HFIP with CF3COONa”.
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
- KD-801 (GPC)
- KD-802 (GPC)
- KD-802.5 (GPC)
- KD-803 (GPC)
- KD-804 (GPC)
- KD-805 (GPC)
- KD-806 (GPC)
- KD-806M (GPC)
- KD-807 (GPC)
- KF-801 (GPC)
- KF-802 (GPC)
- KF-802.5 (GPC)
- KF-803 (GPC)
- KF-803L (GPC)
- KF-804 (GPC)
- KF-804L (GPC)
- KF-805 (GPC)
- KF-805L (GPC)
- KF-806L (GPC)
- KF-806M (GPC)
- KF-807L (GPC)
- KF-401HQ (GPC)
- KF-402.5HQ (GPC)
- KF-402HQ (GPC)
- KF-403HQ (GPC)
- LF-804 (GPC)
- LF-404 (GPC)
- LF-604 (GPC)
Product Name Index
- KD-801 (GPC)
- KD-802 (GPC)
- KD-802.5 (GPC)
- KD-803 (GPC)
- KD-804 (GPC)
- KD-805 (GPC)
- KD-806 (GPC)
- KD-806M (GPC)
- KD-807 (GPC)
- KF-801 (GPC)
- KF-802 (GPC)
- KF-802.5 (GPC)
- KF-803 (GPC)
- KF-803L (GPC)
- KF-804 (GPC)
- KF-804L (GPC)
- KF-805 (GPC)
- KF-805L (GPC)
- KF-806L (GPC)
- KF-806M (GPC)
- KF-807L (GPC)
- KF-401HQ (GPC)
- KF-402.5HQ (GPC)
- KF-402HQ (GPC)
- KF-403HQ (GPC)
- LF-804 (GPC)
- LF-404 (GPC)
- LF-604 (GPC)
Related Products
- Organic SEC (GPC) Columns (General Analysis): THF
- Organic SEC (GPC) Columns (General Analysis): DMF
- Organic SEC (GPC) Columns: High Performance Analysis
- Organic SEC (GPC) Column: Ultra-rapid Analysis
- Organic SEC (GPC) Columns: Linear Calibration Type
- Solvent-peak Separation Columns for Organic SEC (GPC)
- Organic SEC (GPC) Column: Rapid Preparative Column
- Organic SEC (GPC) Columns: Preparative
- Organic SEC (GPC) Columns: Preparative Columns [Customized Columns]