The Japanese Industrial Standards (JIS) K0450-70-10 states the “Testing methods for perfluorooctanesulfonate (PFOS) and perfluorooctanoate (PFOA) in industrial water and wastewater”. It writes quantitative analysis of PFOS and PFOA linear isomers, which are some of the most common PFAS (perfluoroalkyl compounds and polyfluoroalkyl compounds)
Its Annex JC, provides reference analysis conditions for organic fluoride compounds (PFAS) with different carbon chain numbers.
In this application, RSpak JJ-50 2D, a multimode (reversed phase + anion exchange) column was used to analyze four PFAS compounds. The analytes were dissolved in a mixture of water and methanol (1:1). The analytical condition used was referred to the Annex JC, with slight modifications.
*PFOS and PFOA are also often called “perfluorooctane sulfonic acid” and “perfluorooctanoic acid”, respectively. At Shodex we use perfluorooctane sulfonic acid and perfluorooctanoic acid throughout our website.
Sample: 10 µL
10 ng/mL each (in CH3OH/H2O=1/1)
- PFBS;Perfluorobutanesulfonic acid
- PFOS;Perfluorooctanesulfonic acid
- PFBA;Perfluorobutanic acid
- PFOA;Perfluorooctanoic acid
- Column
- :Shodex RSpak JJ-50 2D (2.0 mm I.D. × 150 mm)
- Eluent
- :50 mM CH3COONH4 aq. (adjusted pH 9 with NH3 aq.)/CH3OH=20/80
- Flow rate
- :0.2 mL/min
- Detector
- :ESI-MS (negative, SIM)
- Column temp.
- :40 ℃
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Analysis of Haloacetic Acids (JJ-50 2D)
- LC/MS Analysis of Haloacetic Acids (JJ-50 2D)
- Melamine (1) (NH2P-50 4E)
- Melamine (2) (NH2P-50 2D)
- LC/MS Analysis of Melamine and Cyanuric Acid (NH2P-40 2D)
- Melamine and Related Substances (VG-50 4D)
- Effects of Eluent Composition Ratios on PFOS and PFOA Analysis (JJ-50 2D)
- Analysis of PFAS (VT-50 2D)
- LC/MS Analysis of Organic Arsenic Compound (SI-35 2B)