Analysis of Melamine (2) (NH2P-50 2D)

Asahipak NH2P-50 2D, a polymer-based amino column was used for melamine LC/MS analysis. LC/MS provides high selectivity and high sensitive measurement.

Sample preparation

 

  1.  

Sample: 1 µL
Added 20 µg/L of melamine to the test samples
(Left) Milk
(Center) Powder milk
(Right) Dairy product

  1.  
Column
Shodex Asahipak NH2P-50 2D (2.0 mm I.D. x 150 mm)
Eluent
(A); 5 mM Ammonium formate (pH3 adjusted by formic acid)
(B); CH3CN
Gradient program; 90 % B (0 - 10 min) - 20 % B (10.01 - 20 min) - 90 % B (20.01 - 35 min)
Flow rate
0.2 mL/min
Detector
ESI-MS (Positive m/z=127.1)
Column temp.
40 ℃

Data provided by Shimadzu Corporation.