LC/MS Analysis of Melamine and Cyanuric Acid (NH2P-40 2D)

Cyanuric acid is a structural analogue of melamine, and thus sometimes detected as an impurity of melamine. In this application, polymer-based amino column, Asahipak NH2P-40 2D, was used to separate melamine and cyanuric acid using HILIC mode. Hydrophilic compounds are retained at a high concentration acetonitrile under HILIC mode condition. This is advantageous for the high sensitivity analysis by ESI-MS.
*Sales of Asahipak NH2P-40 2D used in this application is discontinued. Please consider trying Asahipak NH2P-50 2D.

Sample: 5 μL
500 ng/mL each

  1. 1.Melamine
  2. 2.Cyanuric acid
Column
Shodex Asahipak NH2P-40 2D (2.0 mm I.D. x 150 mm)
Eluent
50 mM HCOONH4 aq./CH3CN=25/75
Flow rate
0.2 mL/min
Detector
(Melamine) ESI-MS/MS, (Cyanuric acid) ESI-MS
Column temp.
30 ℃