Inorganic ions are generally monitored using an electrical conductivity detector. However, inorganic ions that have UV absorption, such as nitrite and nitrate ions, can also be detected using a UV detector. Thus, they can be analyzed without using an ion chromatography system nor an ion chromatography column. In this application, nitrite and nitrate ions were separated using a polymer-based amino column Asahipak NH2P-50 4E.
Sample: 10 µL, 10 mg/L each
- 1.NO2-
- 2.NO3-
- Column
- :Shodex Asahipak NH2P-50 4E (4.6 mm I.D. x 250 mm)
- Eluent
- :1.79 g Na2HPO4 ・ 12 H2O + 0.78 g NaH2PO4 ・ 2 H2O + 14.04 g NaClO4 ・ H2O in 1 L H2O
- Flow rate
- :0.8 mL/min
- Detector
- :UV (210 nm)
- Column temp.
- :40 ℃