Samples containing more than 10 mg/L of Mg2+ and/or 20 mg/L of Ca2+ must be pretreated, because those divalent cations form complexes with the eluent and give rise to broad peaks which interfere with the analysis. For pretreatment, samples are passed through a pretreatment cartridge packed Na+ type cation exchange resin. This replaces the divalent cations in samples by Na+ and removes the interfering peaks.
(Pretreatment with the cartridges packing Na+ type cation exchange resin)
1. Rinse the cartridge with 3 mL of methanol and 3 mL of H2O separately.
2. Load 3 mL of the sample onto the cartridge.
3. Waste 1 mL of the first elution.
4. Collect 2 mL of following elution.
Sample : Mineral water, 100 μL
1. HCO3-
2. Cl-
3. NO3-
4. SO42-
Column : Shodex IC SI-90 4E (4.0 mm I.D. x 250 mm) Eluent : 5 mM p-Hydroxybenzoic acid + 5.3 mM N,N-Diethylethanolamine aq. Flow rate : 1.5 mL/min Detector : Non-suppressed conductivity Column temp. : 25 °C