Repeatability for ODP-50 4E, A Polymer-based Column for Reversed Phase Chromatography, under Alkaline Conditions

When testing the repeatability of Asahipak ODP-50 4E, a polymer-based column for reversed phase chromatography, under alkaline condition, azithromycin was injected repeatedly according toUSP 39 method. When theoretical plate number for azithromycin and resolution of azithromycin/azaerythromycin A regarding number of injections were measured, theoretical plate number and resolution were not changed after 3,000 injections. It was confirmed that ODP-50 4E is very alkaline-resistant.

Sample: 10 μL
0.5 mg/mL each

  1. 1.Azaerythromycin A
  2. 2.Azithromycin

dissolved in 6.7 g/L Dibasic potassium phosphate aq. (pH8.0 adjusted with phosphoric acid)/CH3CN=40/60


Column
Shodex Asahipak ODP-50 4E (4.6mmI.D. x 250mm)
Eluent
6.7 g/L Dibasic potassium phosphate aq. (pH11.0 adjusted with 10 M KOH)/CH3CN=40/60
Flow rate
1.0 mL/min
Detector
UV (210 nm)
Column temp.
40 ℃

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Related Products

Product Information (Related Information)