The relation between column temperature and retention time and the relation between column temperature and theoretical plate number (TPN) for ODP2 HP 4D (4.6 mm I.D. x 150 mm length) are shown. As retention time and TPN vary with changes in column temperature, the use of a column oven for temperature control is recommended.
Sample: 5 μL
- Phenol 300 ppm
- Methyl benzoate 350 ppm
- Toluene 1000 ppm

* Ratio of TPN [%] : TPN at 40 ℃ is set as 100%.

- Column
- :Shodex ODP2 HP-4D (4.6 mm I.D. x 150 mm)
- Eluent
- :H2O/CH3CN=55/45
- Flow rate
- :0.5 mL/min
- Detector
- :UV (254 nm)