Effects of Sample Injection Volume (ODP2 HP)

The relation between sample injection volume and theoretical plate number (TPN) for ODP2 HP 4D (4.6 mm I.D. x 150 mm length) is shown. Sample injection volume of 40 μL and below is recommended for ODP2 HP-4D to achieve maximum efficiency. Sample injection volume of 8 μL and below is recommended for ODP2 HP-2D (2 mm I.D. x 150 mm length).

Sample: 0.6 µg each

  1. Phenol
  2. Methyl benzoate
  3. Caffeine

* Ratio of TPN [%] : TPN at sample injection of 2 μL is set as 100%.

Column
Shodex ODP2 HP-4D (4.6 mm I.D. x 150 mm)
Eluent
H2O/CH3CN=55/45
Flow rate
1.0 mL/min
Detector
UV (254 nm) Phenol, Methyl benzoate
UV (300 nm) Caffeine
Column temp.
40 ℃