Five kinds of monoamine neurotransmitters were analyzed using HILICpak VC-50 2D, a polymer based column for HILIC mode. Histamine is more basic than other monoamines and causing a stronger ion exchange interactions between histamine and gel material resulting histamine to elute later.
Sample :
0.1 μM each (in H2O), 20 μL
Noradrenaline
Adrenaline
Dopamine
Serotonin
Histamine
Column : Shodex HILICpak VC-50 2D (2.0 mm I.D. x 150 mm) Eluent : (A) 200 mM HCOOH aq. (B) CH3CN Linear gradient (High pressure) ; B % = 60 % (0 to 5 min) 60 % to 10 % (5 to 6 min) 10 % (6 to 20 min) Flow rate : 0.3 mL/min Detector : ESI-MS/MS (MRM Positive) Column temp. : 40 ℃